Apera Instruments 2310-C conductivity electrode is designed for high range conductivity measurement (20 to 2,000 mS/cm, K=10). The electrode is equipped with platinum black coated ring sensor and provides stable and accurate high range conductivity measurement.
The accuracy is ±1.5 F.S. Even without calibration, the accuracy can be maintained at ±10% of full scale.
Compatible with all conductivity meters with BNC connectors. The temperature sensor is not installed. For full temperature compensation, a separate temperature probe is required to activate ATC or enter the temperature manually.
The electrode has a 6-month manufacturer’s warranty.